COMPARISON · MULTIMETERS

Fluke 289 vs Fluke 115: Is a Logging Multimeter Worth It?

A logging multimeter like the Fluke 289 is worth it if you regularly chase faults that come and go when nobody is watching, or you have to hand someone a record of what a circuit did. If you mostly find a fault and fix it on the spot inside CAT III 600V, the Fluke 115 does that job and the 289's extra cost buys features you won't use. Both are True RMS Flukes; the 289 adds time-stamped recording, a microamp range, CAT IV 600V, temperature and 50,000 counts, while the 115 is lighter and simpler.

By the OhmSavvy Research Team ·

Disclosure: OhmSavvy earns a commission from qualifying Amazon purchases made through these links, at no extra cost to you. How we're funded.

Fluke 289 True RMS logging multimeter with TrendCapture, front view
Best for Low Resistance

Fluke 289

4.7★★★★★

The only handheld here with a 50 Ω range at 1 mΩ resolution, plus 0.01 µA current resolution and on-screen trend logging.

Fluke 115 True RMS field-service multimeter, front view
Best for Field ServiceHands-on Tested

Fluke 115

4.8★★★★★

The same Fluke reliability as the 117, in the same body, without the electrician-specific extras (NCV, AutoVolt, LoZ).

Specs side by side

Fluke 289Fluke 115
Rating4.7★★★★★4.8★★★★★
AccuracyTrue RMS, 0.025% basic DC VTrue RMS
Safety ratingCAT IV 600V / CAT III 1000VCAT III 600V
Auto-rangingYes (manual override) Yes
Best forLow-ohm, µA and logged electronics workField technicians, general electrical and electronics work
RecordingUp to 10,000 intervals plus events, real-time clock, on-screen graphMin/Max/Avg with elapsed time
PC transferIR link, FlukeView FormsNone listed
Display50,000 counts6,000 counts + 33-segment bar graph
DC volts accuracy±(0.025% + 2)±(0.5% + 2)
AC volts accuracy±(0.3% + 25) at 45–65 Hz±(1.0% + 3) at 45–500 Hz
Current500 µA (0.01 µA) to 10 A6 A and 10 A only; 1 mA best resolution
Resistance50 Ω range at 0.001 Ω; to 500 MΩTo 40 MΩ
CapacitanceTo 100 mF1 nF to 9,999 µF
FrequencyTo 999.99 kHzTo 50 kHz (2020 manual: 99.99 kHz)
Temperature-200 to 1350 °C, excluding probeNo (Fluke 116 only)
LoZ / low-pass filterYes / yes (above 1 kHz)No / no
Safety ratingCAT III 1000V / CAT IV 600VCAT III 600V
Battery6 AA; 100 h minimum, 200 h loggingOne 9 V, about 400 h
Weight871 g550 g
WarrantyLifetime limited3 years
OhmSavvy statusDesk research against Fluke's manualHands-on Tested

Specs checked against manufacturer documentation and official Amazon listings. Fluke 115 is hands-on tested by OhmSavvy; the Fluke 289 assessment is desk research. How we research.

Key differences

  • →Logging: the 289 records up to 10,000 intervals plus events against a real-time clock and graphs them on screen. The 115 keeps one minimum, one maximum and an average, with elapsed time.
  • →Documentation: the 289's IR link sends recordings to FlukeView Forms on a PC; Fluke lists no PC link for the 115.
  • →Safety: the 289 is rated CAT IV 600V / CAT III 1000V; the 115 stops at CAT III 600V.
  • →Low current: the 289 resolves 0.01 µA; the 115 has no mA or µA input and resolves 1 mA at best.
  • →Extras only on the 289: temperature, LoZ, a 1 kHz low-pass filter, a 50 Ω low-ohm range and 250 µs peak capture.
  • →Living with it: the 115 is 550 g and runs about 400 hours on one 9 V battery; the 289 is 871 g on six AA cells, rated 100 hours minimum.
  • →Shared: both are True RMS, auto-ranging Flukes. The 289's warranty is lifetime limited; the 115's is 3 years.

What logging actually buys you

"Logging" on the Fluke 289 means three things working together. The meter stores readings as intervals and events, up to 10,000 intervals in Fluke's manual. Each record carries a time from the built-in real-time clock. And TrendCapture draws the record as a graph on the meter's own screen, so you can review a dip without a laptop. The IR link then moves the data to FlukeView Forms when you need a printable report.

The Fluke 115 has a simpler tool for the same problem: Min/Max/Avg with elapsed time. Leave it on a circuit and it keeps the lowest reading, the highest and the average, and tells you how long after you started each extreme occurred. That's genuinely useful, and it's more than many meters offer. But it holds one minimum. If the fault happens twice, the shallower event disappears.

120 V109 V23:4104:12Fluke 115 Min/Max/AvgMIN 109 V, elapsed 6:12first dip is overwrittenFluke 289 recording23:41 112 V, 04:12 109 Vboth events, clock timesIllustrative values, not a recording.
Illustrative values. Min/Max/Avg keeps the worst dip; a recording keeps every dip with its clock time.

The difference matters most with repetition. Two dips at 23:41 and 04:12 that both match a well pump or a furnace cycle point to a cause. A single minimum with an elapsed time of six hours confirms a problem and stops there. Our guide to finding an intermittent electrical fault walks through using either approach.

How long a recording can run

Two limits decide how long the 289 can watch a circuit: memory and battery. Fluke's manual gives up to 10,000 recorded intervals and a battery life of 200 hours in logging mode. Our data doesn't list the interval settings, so here is illustrative arithmetic only: if the meter closed one interval per minute, 10,000 intervals would cover 10,000 ÷ 60 ≈ 167 hours, about 6.9 days, and the 200-hour battery figure (about 8.3 days) would outlast it. At one interval every 30 seconds, memory fills in about 83 hours, roughly three and a half days.

1 per 30 s
≈83 h
1 per minute
≈167 h
Battery, logging
200 h
Illustrative coverage of 10,000 intervals at two assumed rates, against the 200 h logging battery figure.

Either way, a weekend or a full work week is within reach, which is the window most intermittent complaints need. The 115 has no memory limit to worry about because it stores only three values, and its 9 V battery is rated about 400 hours, so it can sit in Min/Max/Avg for a long time too. It just can't tell you much when you come back.

Intermittent faults: which meter solves which

ComplaintFluke 115 approachFluke 289 approach
Lights dim a few times a nightMin/Max shows the worst dipEach dip with its time
Breaker trips once a weekMin/Max keeps the peak load ampsEach current peak with its time
Furnace drops flame at randomNo µA range for the flame rod0.01 µA steps on the 500 µA range
Equipment resets on a scheduleElapsed time of one eventPattern across many events
Customer disputes the diagnosisYour word and a photo of the displayExported record in FlukeView Forms
Fault happens while you watchBar graph and live readingLive reading, slower to set up

The last row is the honest one. When the fault is present and steady, the 115 is quicker: dial, probe, read. The 289's strength is the fault that won't perform on demand. For voltage complaints in a home, our guide to checking for voltage fluctuations shows what to look for in a recording.

Precision and safety, beyond logging

Logging isn't the only thing the 289's higher tier buys. Its DC accuracy is twenty times tighter on the percentage term, its display has more than eight times the counts, and its rating extends to CAT IV 600V. Here is what the accuracy gap means at two readings, using each meter's published line and the resolution of the range the reading lands on.

ReadingFluke 289 allowanceFluke 115 allowance
24.000 V DC control circuit0.006 + 0.002 = ±0.008 V0.12 + 0.02 = ±0.14 V
120.0 V AC outlet, 60 Hz0.36 + 0.25 = ±0.61 V1.2 + 0.3 = ±1.5 V
Fluke 289
±0.025%
Fluke 115
±0.5%
Basic DC volts accuracy, percentage term only. Lower is tighter.
Fluke 289
0.01 µA
Fluke 115
1 mA

Bars use a logarithmic scale: each step is ten times the last.

Finest current step on a log scale. The 115 has no mA or µA input.

The arithmetic: 0.5% of 24 V is 0.12 V, plus 2 counts of 0.01 V on the 115's 60.00 V range; the 289 adds 2 counts of 0.001 V to its 0.006 V. At 120 V AC the 115 sits on its 600.0 V range and the 289 on 500.00 V. These are worst-case limits from published specs, not measurements. For checking an outlet or a 24 V thermostat circuit, ±1.5 V and ±0.14 V are entirely adequate, which is why the 115 is enough for most field service.

CAT III 600V is a real boundary

The 115 is rated for branch circuits, receptacles and distribution panels inside the building. It isn't rated for the service entrance, meter base or outdoor service conductors, where CAT IV applies; the 289 is. Whichever meter you use, prove it on a known live source before and after a dead test, and leave utility-side work to the utility or a licensed electrician.

So, is a logging multimeter worth it?

The 289 sits in Fluke's top handheld tier and the 115 in its mid-range field tier, and its price reflects that. We don't quote prices because they change; the decision rests on how often you would use what the extra money buys.

Worth it

You get callbacks for faults you can't reproduce, and every repeat visit costs you a day.

Worth it

You hand over records: commissioning, warranty claims, landlord or utility disputes.

Worth it

You also need µA, temperature or CAT IV, which the 115 can't cover at all.

Not worth it

You find faults live and fix them on the visit, inside CAT III 600V.

Not worth it

You'd log a few times a year; a dedicated logger or a borrowed meter covers that.

Not worth it

Weight and battery swaps matter more to you than features you rarely open.

  1. 01

    Does the fault show while you watch?

    Yes: the 115 is enough

  2. 02

    Need µA, temperature or CAT IV?

    Yes: the 115 can't do it

  3. 03

    Do you hand over records?

    Yes: the 289 earns its cost

  4. 04

    Otherwise

    Buy the 115; borrow a logger

What our hands-on time with the 115 tells us

OhmSavvy bought and used a Fluke 115, which is why it carries our Hands-on Tested badge. We haven't published bench measurements, and nothing in our use contradicted Fluke's specifications. We haven't tested the 289, so its column rests on Fluke's manual. Read the Fluke 115 review for more.

Which should you buy?

You are…Better fitWhy
Appliance or field-service techFluke 115True RMS, 10 A, capacitance, light
Residential electrician on callbacksFluke 289Overnight record of the problem
HVAC tech checking flame rodsFluke 2890.01 µA steps; the 115 has no µA
Facilities tech, verify-and-fix roundsFluke 115Fast readings inside CAT III 600V
Controls or commissioning engineerFluke 289Records and FlukeView Forms export
Work at the meter base or service dropFluke 289CAT IV 600V rating
Homeowner or first serious meterFluke 115Covers household work without extras

For an industrial take on the 289, see Fluke 289 vs Fluke 179; for the 115 against a bigger field meter, see Fluke 115 vs Fluke 87V. If logging appeals but your work is bench electronics, compare Fluke 289 vs Fluke 287.

Choose the Fluke 289 if…

  • ✓You need to leave a meter on a circuit overnight and see every event with its clock time
  • ✓You write up reports, warranty claims or customer evidence from recorded readings
  • ✓You also need µA, temperature, CAT IV 600V or a drive low-pass filter

Choose the Fluke 115 if…

  • ✓You diagnose and fix on the spot: outlets, appliances, controls and branch circuits up to 600 V
  • ✓Min/Max/Avg with elapsed time is enough to tell you a dip happened
  • ✓You want a lighter meter we've hands-on tested, with about 400 hours on one 9 V battery

The bottom line

A logging multimeter is worth it when faults hide from you or someone needs proof: the Fluke 289 records every event with a clock time, and adds µA, temperature and CAT IV 600V. If you mostly find and fix faults on the visit inside CAT III 600V, the Fluke 115 is the better buy: True RMS, lighter, about 400 hours per battery, and a meter we've hands-on tested.

Frequently asked questions

Can the Fluke 115 log data?+

Not in the sense the 289 does. The 115 has Min/Max/Avg with elapsed time, which keeps the lowest, highest and average readings and how long after the start each extreme happened. It has no memory of separate events, no real-time clock and no PC link listed. For a timeline of every dip or surge you need a logging meter such as the 289.

What is TrendCapture on the Fluke 289?+

TrendCapture is Fluke's name for the 289's on-screen graph of recorded readings. After a recording session you can view the trend on the meter itself without a computer. The recording holds up to 10,000 intervals plus events against a real-time clock, and the IR link sends it to FlukeView Forms when you need a report.

How long can the Fluke 289 record?+

It depends on the interval setting, which our data doesn't list. Fluke's manual gives up to 10,000 recorded intervals and 200 hours of battery in logging mode. As illustrative arithmetic, one interval per minute would fill memory in about 167 hours, close to seven days, before the batteries run down. Start long runs with fresh AA cells.

Is the Fluke 289 overkill for home electrical work?+

Usually, yes. Household outlets, switches, appliances and branch circuits sit inside the 115's CAT III 600V rating, and its True RMS and 6,000-count display handle them well. The 289 earns its place at home only for a fault that comes and goes, such as nightly dimming, where a recording shows when it happens and how often.

Does the Fluke 115 measure temperature or microamps?+

No to both. Fluke's datasheet gives the thermocouple input to the Fluke 116, not the 115, and the 115's current inputs are 6 A and 10 A with 1 mA best resolution. The 289 reads temperature from -200 to 1350 °C excluding the probe and resolves 0.01 µA on its 500 µA range, which covers HVAC flame-rod checks.

Is a dedicated data logger better than a logging multimeter?+

For permanent or multi-week monitoring of one quantity, a dedicated logger is often simpler. The 289's advantage is that the same meter does live troubleshooting and recording, so you can switch from probing to logging on the spot. If you rarely log, a 115 for daily work plus a borrowed or rented logger can cover the occasional long job.