COMPARISON · MULTIMETERS

Fluke 116 vs Fluke 115 for HVAC Troubleshooting

For HVAC troubleshooting the Fluke 116 is the better meter: it reads flame-rod current on a 600.0 µA range with 0.1 µA resolution, takes Type K temperature with the included 80BK probe, and has a LoZ input for ghost voltage on 24 V control wiring. The Fluke 115 has none of those three. What the 115 keeps is 6 A and 10 A current ranges. On capacitors, line voltage and resistance, the two read identically.

By the OhmSavvy Research Team ·

Disclosure: OhmSavvy earns a commission from qualifying Amazon purchases made through these links, at no extra cost to you. How we're funded.

Fluke 116 HVAC multimeter with microamp and temperature functions, front view
Best HVAC Multimeter

Fluke 116

4.6★★★★★

Fluke's HVAC multimeter: a 0.1 µA flame-sensor range, temperature, capacitance and LoZ in a CAT III 600V body.

Fluke 115 field service True RMS multimeter, front view
Best for Field ServiceHands-on Tested

Fluke 115

4.8★★★★★

The same Fluke reliability as the 117, in the same body, without the electrician-specific extras (NCV, AutoVolt, LoZ).

Specs side by side

Fluke 116Fluke 115
Rating4.6★★★★★4.8★★★★★
AccuracyTrue RMSTrue RMS
Safety ratingCAT III 600VCAT III 600V
Auto-rangingNot stated by Fluke (manual RANGE button) Yes
Best forHVAC flame sensors, capacitors, controlsField technicians, general electrical and electronics work
Microamps (flame sensor)600.0 µA AC/DC, 0.1 µA steps No
Amps No6.000 A and 10.00 A AC/DC
Temperature-40 to 400 °C / -40 to 752 °F, Type KNo (see the note on Fluke's 115 page below)
Temperature probe includedYes, 80BK No
AutoVolt with LoZ inputYes, about 3 kΩ No
DC millivolts600.0 mV, ±(0.5% + 2)600.0 mV, ±(0.5% + 2)
Capacitance1 nF to 9,999 µF, ±(1.9% + 2)1 nF to 9,999 µF, ±(1.9% + 2)
Resistance / continuityTo 40.00 MΩ; beep below 20 ΩTo 40.00 MΩ; beep below 20 Ω
Safety ratingCAT III 600VCAT III 600V
Size / weight167 × 84 × 46 mm, 550 g167 × 84 × 46 mm, 550 g
Warranty3 years3 years

Specs checked against manufacturer documentation and official Amazon listings. Fluke 115 is hands-on tested by OhmSavvy; the Fluke 116 assessment is desk research. How we research.

Key differences

  • →Only the 116 reads microamps: 600.0 µA AC or DC at 0.1 µA resolution, the range Fluke built for flame sensors. The 115's finest current step is 1 mA on its 6.000 A range.
  • →Only the 116 measures temperature, with a Type K input from -40 to 400 °C and an 80BK probe in the box. Fluke's datasheet and manual list no temperature function on the 115.
  • →Only the 116 has AutoVolt with a low-impedance input of about 3 kΩ. Every volts position on the 115 stays high impedance, so it can show ghost voltage on idle control conductors.
  • →Only the 115 reads amps through the leads, on 6.000 A and 10.00 A ranges. The 116 has no amp or milliamp range.
  • →Capacitance, line voltage, millivolts, resistance, frequency, the case, battery and 3-year warranty are the same on both.

Fluke 116 vs 115 spec by spec for HVAC jobs

Rather than list every range, the table below takes the functions an HVAC technician actually reaches for on a service call and reads each meter's line from Fluke's specification tab.

HVAC functionFluke 116Fluke 115
Flame-sensor current600.0 µA, DC ±(1.0% + 2)Not resolvable (1 mA steps)
Air or pipe temperatureType K, ±(1% + 10) at 0.1 °CNot listed in the datasheet
Pilot thermocouple millivolts600.0 mV DC, ±(0.5% + 2)600.0 mV DC, ±(0.5% + 2)
Run or start capacitorTo 9,999 µF, ±(1.9% + 2)To 9,999 µF, ±(1.9% + 2)
24 V control voltageV AC or AutoVolt LoZV AC (high impedance only)
Line voltage, 120–480 VTo 600.0 V, True RMSTo 600.0 V, True RMS
Motor winding, igniter ohmsTo 40.00 MΩTo 40.00 MΩ
Control-circuit currentOnly to 600 µA6.000 A range, 1 mA steps
Blower or compressor ampsClamp meter neededClamp meter in practice
Fluke 116 (µA range)
0.1 µA
Fluke 115 (6 A range)
1,000 µA

Bars use a logarithmic scale: each step is ten times the last.

Finest current step on each meter, in microamps, log scale. A flame signal of a few µA disappears below the 115's 1 mA step.
115, V AC
>5 MΩ
116, V AC
>5 MΩ
116, AutoVolt LoZ
≈3 kΩ

Bars use a logarithmic scale: each step is ten times the last.

Volts input impedance, log scale. The 116's AutoVolt LoZ input loads idle conductors enough to drain coupled voltage.

Fluke's 115 page lists temperature; its datasheet doesn't

The comparison table on Fluke's 115 product page shows a temperature range for the 115. The 110 Series datasheet, the users manual and the 115's own accuracy table list temperature for the 116 only, and the 115 has no 80BK probe in its box. We follow the datasheet and manual: don't buy the 115 to read temperature.

Flame-rod microamps: the 116's reason to exist

A gas furnace proves flame by flame rectification. The control board puts AC on the flame rod; the flame conducts better in one direction than the other, so a small DC current flows from rod to the grounded burner. If that current is too weak, the board shuts the gas valve a few seconds after ignition. The classic symptom is a burner that lights, runs briefly and drops out, then retries until lockout. The only way to measure the signal is a meter on a microamp range in series with the rod wire.

Control boardAC on theflame-sense terminalµAmeter in seriesflame rodburner (grounded)flame conductsmostly one waychassis ground back to the board
Flame rectification and where the meter goes. Use the DC µA function; the current is tiny but the circuit is live.

Flame signals are commonly a few microamps; the minimum each control accepts is set by its maker and is usually printed in the furnace manual. Work the arithmetic on an illustrative 3.0 µA reading. The 116's DC µA line allows 1.0% of 3.0 µA (0.03 µA) plus 2 counts of 0.1 µA, so about ±0.23 µA: easily good enough to tell a healthy signal from a marginal one. The 115 on its 6.000 A range shows 0.000 A for the same current, because 3 µA is a third of a percent of its smallest step.

3.0µA DCµA (600.0 range)
Illustrative: the 116 in series with a flame rod.
0.000A DCA (6.000 range)
Illustrative: the 115 on the same circuit.

A low reading usually points to an oxidized rod, a poor burner ground or a cracked insulator rather than the board. Our guide to measuring microamps with a multimeter covers the series hookup in detail.

Temperature and thermocouples

Temperature comes up on almost every HVAC call: supply and return air for a temperature split, a line temperature for superheat or subcooling, a heat-exchanger outlet. The 116 does it with the bundled 80BK probe. Fluke's accuracy is ±(1% + 10) on the 0.1 °C scale, so 10 counts add 1.0 °C; at 13 °C supply air that is about ±1.1 °C, before probe error, which Fluke excludes from the figure. In Fahrenheit the line is ±(1% + 18) at 0.2 °F steps, about ±4.2 °F at 55 °F.

That is fine for a temperature-split check. A rule of thumb many technicians use for cooling is a split around 15 to 22 °F, but the equipment maker's chart wins. Taking both readings with the same probe and meter helps, because a fixed offset largely cancels when you subtract one reading from the other.

The pilot thermocouple on a standing-pilot appliance is a different measurement, and here the two meters tie. It is a millivolt reading, not a temperature reading, and both meters have the same 600.0 mV DC range at ±(0.5% + 2). Open-circuit values in the 20s of millivolts are commonly quoted for a healthy pilot thermocouple; check the valve maker's figure. On an illustrative 25.0 mV reading, either meter is good to about ±0.3 mV. Our guide to checking a thermocouple with a multimeter walks through it.

13 °C supply air
±1.1 °C
24 °C return air
±1.2 °C
60 °C heat outlet
±1.6 °C
Allowed error, in °C, on the 116 at three temperatures from its ±(1% + 10) line. Probe error is extra.

Capacitors and LoZ on 24 V controls

Run capacitors are the most common failed part on a condenser, and on this job the meters are equal: both read 1 nF to 9,999 µF at ±(1.9% + 2). On an illustrative 45 µF compressor section, read on the 100.0 µF range, the allowed error is 0.86 µF plus 0.2 µF, about ±1.1 µF. On a 5 µF fan section, read on the 10.00 µF range, it is about ±0.12 µF. Either is tight enough to compare against the tolerance printed on the can. See how to test a fan motor capacitor for the procedure.

Control wiring is where LoZ pays off. Thermostat cable often has spare conductors, and long runs sit next to powered ones, so an unused or open wire can pick up a coupled voltage that a high-impedance input displays as a believable number. On the 115, every volts position stays above 5 MΩ, so you must load-test a suspicious reading yourself. The 116's AutoVolt position drops the input to about 3 kΩ, which drains coupling and leaves a real 24 V supply standing.

  1. 01

    Odd voltage on a W or Y wire

    Call for heat or cooling is off

  2. 02

    116: switch to AutoVolt

    Input drops to about 3 kΩ

  3. 03

    Reading collapses?

    Coupled ghost voltage, not a fault

  4. 04

    Reading holds?

    Real voltage: check thermostat or wiring

Our guides to testing a thermostat and checking a 24 V transformer cover the control circuit step by step.

Where the 115 is the better HVAC meter

The 115's amp ranges are not useless on HVAC work. Setting the heat anticipator on an older mechanical thermostat calls for the current in the heat circuit, which is a fraction of an amp; the 115 reads it in series on its 6.000 A range, and the 116's 600 µA ceiling can't. Small inducer or damper motors and 24 V control loads can also be read in series. Blower, compressor and electric-heat amps are a different story: they are big, often above 10 A, and sit inside running equipment, so a clamp meter is the right tool with either Fluke.

A technician who works mainly on heat pumps, mini-splits and electric heat, with no gas appliances, never needs a flame-sensor range. If that technician already owns an HVAC clamp with a temperature input, the 115 covers the remaining multimeter jobs and adds amps. Our Fluke 115 review is based on the owner's own unit; the 116 assessment here is desk research against Fluke's published specifications.

HVAC specialist or general-purpose meter?

Fluke built the 116 as a specialist and the 115 as a field-service all-rounder, and the split shows the moment the meter leaves the mechanical room. Off an HVAC call, nobody needs flame-rod microamps, but plenty of jobs need amps: a parked car's battery drain, a bench appliance, a small motor. The 116 has no input for any of them, because its only current range tops out at 600.0 µA. Everything voltage-, resistance- and capacitance-based is identical, so the general-purpose gap is current and nothing else.

General jobFluke 116Fluke 115Worked tolerance
12.60 V car battery at rest60.00 V DC range60.00 V DC range±0.08 V on both, from ±(0.5% + 2)
120 V receptacleTrue RMS, 600.0 VTrue RMS, 600.0 V±1.5 V on both, from ±(1.0% + 3)
Parked-car drain, illustrative 50 mANot possible (µA input only)6.000 A range, 1 mA steps115: ±3.5 mA, from ±(1.0% + 3)
Appliance load, illustrative 5 A ACNot possible6.000 A range115: about ±0.08 A, from ±(1.5% + 3)
Fuse, cord or switch continuityBeeps below 20 ΩBeeps below 20 ΩSame
Electrolytic in a power supplyTo 9,999 µFTo 9,999 µFSame ±(1.9% + 2)
Idle conductor showing phantom voltsAutoVolt LoZHigh impedance only116 can tell coupling from a fault
Oven or freezer temperatureType K, 80BK probeNot listed116 only
Fluke 116
7 of 9
Fluke 115
6 of 9
Rows in the HVAC table near the top of this page each meter handles itself (9 rows; clamp-only jobs count against both).
Fluke 116
6 of 8
Fluke 115
6 of 8
Rows in the general-job table above each meter handles itself (8 rows).

The tie in the second chart hides a difference in kind. The two rows the 116 loses are current, which comes up in car, appliance and electronics work every week. The two the 115 loses, LoZ and temperature, are handy but have workarounds: a load across the leads, or a separate thermometer. So a homeowner, a car hobbyist or a general maintenance tech who rarely sees a gas burner gets more from the 115, and our owner-tested Fluke 115 review covers it as an everyday meter. The 116 only becomes the better general meter if you already own a clamp for amps. If you want both amps and LoZ in this body, look at the 117 instead; the Fluke 115 vs Fluke 117 page weighs that step.

Fluke 116 vs Fluke 115 for HVAC: which should you buy?

Your HVAC workBetter fitWhy
Residential gas furnaces and boilersFluke 116Flame-rod µA and temperature are routine on these calls
Mixed heating and cooling serviceFluke 116 + clampThe clamp covers amps; the 116 covers everything else
Heat pumps and mini-splits onlyFluke 115 or 116No flame sensors; the 116 still adds temperature and LoZ
Older thermostats with anticipatorsFluke 115Reads heat-circuit current in series; the 116 can't
Refrigeration and superheat workFluke 116Line temperature from the included probe
Already own an HVAC clamp with µAFluke 115Avoids paying twice for µA and temperature
Apprentice building a first kitFluke 116Covers more HVAC readings out of one box

For HVAC clamps and other meters ranked for this trade, see our best multimeter for HVAC technicians guide. If you are deciding between the 116 and the electrician's model, read Fluke 116 vs Fluke 117.

Safety on furnace and condenser work

Shut off power at the disconnect before you open a series path, and reconnect the meter before restoring power for a flame-rod reading; keep your hands clear of the burner and line terminals while the unit runs. Discharge run and start capacitors with a suitable resistor before you test them, because they can hold a dangerous charge after the power is off. Both meters are rated CAT III 600V with a 600 V maximum between any terminal and earth; neither belongs on the utility side of the service. If you smell gas, leave the building and call your gas utility or 911 from outside.

Choose the Fluke 116 if…

  • ✓You service gas furnaces and boilers and need to read flame-sensor microamps
  • ✓You want supply and return air temperature from the meter on your belt
  • ✓You chase phantom voltages on thermostat and control cable runs

Choose the Fluke 115 if…

  • ✓Your HVAC work is mostly cooling, heat pumps and electric heat with no flame sensors
  • ✓You want to read control-circuit or small-motor current in series up to 10 A
  • ✓You already carry an HVAC clamp with µA and temperature functions

The bottom line

For HVAC troubleshooting, buy the Fluke 116: flame-rod microamps, Type K temperature with a probe in the box and a LoZ input for control wiring are the readings that fill service calls. Pair it with a clamp meter for amps. The Fluke 115 suits technicians with no gas equipment who already own an HVAC clamp, or who want series current for anticipators and small loads.

Frequently asked questions

Can the Fluke 115 test a flame sensor?+

Not in practice. Its finest current step is 1 mA on the 6.000 A range, and flame signals are typically a few microamps, so the display stays at zero. The Fluke 116 has a 600.0 µA range with 0.1 µA resolution designed for this check, and several HVAC clamp meters also include a µA range for flame rods.

Does the Fluke 115 measure temperature?+

Fluke's sources disagree. The comparison table on the 115 product page shows a temperature range, but the 110 Series datasheet, the users manual and the 115's accuracy table list temperature only for the 116, and no probe ships with the 115. We follow the datasheet and manual, so treat the 115 as having no temperature function.

Is the Fluke 116 good for HVAC?+

Yes, it is Fluke's dedicated HVAC model. It reads flame-sensor current to 0.1 µA, Type K temperature from -40 to 400 °C with the included 80BK probe, capacitance to 9,999 µF, and has a LoZ input against ghost voltage. Its gap is amps: there is no amp range, so carry a clamp meter for motors and compressors.

Which Fluke is best for testing HVAC capacitors?+

Either. The 116 and 115 share the same capacitance function, 1 nF to 9,999 µF with ±(1.9% + 2) accuracy, which covers run and start capacitors comfortably. A 45 µF reading carries about ±1.1 µF of allowed error. Always discharge the capacitor safely and isolate it from the circuit before you measure it.

Why does my meter show voltage on a thermostat wire that should be off?+

Often it is ghost voltage: an idle conductor beside powered ones in the same cable picks up a coupled voltage that a high-impedance meter displays. A low-impedance input such as the 116's AutoVolt position loads the wire enough to collapse it. The 115 has no LoZ mode, so add a load across the leads to check.

Is the Fluke 116 a good general-purpose multimeter?+

Only if you carry a clamp meter. Its volts, resistance, continuity and capacitance match the Fluke 115, and it adds LoZ and temperature, but its single current range stops at 600.0 µA. That rules out battery-drain, appliance and small-motor current through the leads. For all-round home, car and shop use, the 115 or 117 is the better fit.

Can the Fluke 116 measure amps on a blower motor?+

No. The 116's only current range is 600.0 µA, so it cannot read any motor current. Use a clamp meter around one conductor instead. The 115 can read up to 10 A in series, but breaking into a running blower circuit is awkward and many motors exceed that, so a clamp is the better tool there too.