COMPARISON · MULTIMETERS
Fluke 177 vs Fluke 179: Features and Accuracy Compared
The Fluke 179 is a Fluke 177 with a thermometer: Fluke publishes the same accuracy line for both meters on DC volts (±0.09% + 2), AC volts, AC and DC current, resistance, capacitance and frequency, and only the 179 adds a temperature function, −40 °C to 400 °C at ±(1.0% + 10), with an 80BK probe in the box. Same 420 g body, same CAT IV 600V / CAT III 1000V rating, same backlight, battery and lifetime warranty. Buy the 179 if you will measure temperature; otherwise the 177 gives identical electrical readings.
By the OhmSavvy Research Team ·
Disclosure: OhmSavvy earns a commission from qualifying Amazon purchases made through these links, at no extra cost to you. How we're funded.

Fluke 179
4.7★★★★★Fluke's do-everything bench-and-field meter: CAT IV 600V, True RMS, temperature and a 6,000-count display with bar graph.

Fluke 177
4.6★★★★★The middle 170-series meter: 0.09% DC volts, True RMS to 10 A and a backlight, without paying for the 179's thermometer.
Specs side by side
| Fluke 179 | Fluke 177 | |
|---|---|---|
| Rating | 4.7★★★★★ | 4.6★★★★★ |
| Accuracy | True RMS | True RMS, ±(0.09% + 2) DC V |
| Safety rating | CAT III 1000V / CAT IV 600V | CAT III 1000V / CAT IV 600V |
| Auto-ranging | Yes (manual override) | Yes (manual override) |
| Best for | Industrial maintenance, electronics, HVAC | Everyday electrical troubleshooting |
| Temperature | −40 °C to 400 °C, ±(1.0% + 10), 0.1 °C res | Not available |
| Probe in the box | 80BK temperature probe | None (test leads only) |
| DC volts | ±(0.09% + 2), 0.1 mV res, 1000 V | ±(0.09% + 2), 0.1 mV res, 1000 V |
| AC volts (True RMS) | ±(1.0% + 3), 1000 V | ±(1.0% + 3), 1000 V |
| DC / AC current | ±(1.0% + 3) / ±(1.5% + 3), 0.01 mA to 10 A | ±(1.0% + 3) / ±(1.5% + 3), 0.01 mA to 10 A |
| Resistance | ±(0.9% + 1), to 50 MΩ | ±(0.9% + 1), to 50 MΩ |
| Capacitance | ±(1.2% + 2), to 10,000 µF (9,999 µF in one Fluke table) | ±(1.2% + 2), to 10,000 µF |
| Frequency | ±(0.1% + 1), to 100 kHz | ±(0.1% + 1), to 100 kHz |
| Display | 6,000 counts, analog bar graph, backlight | 6,000 counts, analog bar graph, backlight |
| Battery | 9 V, about 200 h without backlight | 9 V, about 200 h without backlight |
| Size / weight | 43 × 90 × 185 mm, 420 g | 43 × 90 × 185 mm, 420 g |
| Warranty | Lifetime | Lifetime |
Specs checked against manufacturer documentation and official Amazon listings. Desk research, not bench testing. How we research.
Key differences
- →Only the 179 measures temperature: −40 °C to 400 °C at ±(1.0% + 10) with 0.1 °C resolution, using the 80BK probe Fluke includes.
- →Every electrical accuracy line Fluke publishes is the same on both, including the 0.09% DC volts spec.
- →Both share the backlit display with analog bar graph, CAT IV 600V / CAT III 1000V, the 420 g case, about 200 hours on a 9 V cell and a lifetime warranty.
- →Fluke's 179 page contradicts itself in places: its comparison table lists 9,999 µF and a 400 °F top temperature, and names µA as a function, while its spec table gives 10,000 µF, 400 °C and 0.01 mA best current resolution. We follow the spec table.
- →Neither has LoZ, non-contact voltage or a low-pass filter.
Fluke 177 vs 179 at a glance
One extra function
Temperature on the 179, −40 °C to 400 °C, with the 80BK probe included.
Same electrical accuracy
0.09% DC volts, 1.0% AC volts and identical amps, ohms, capacitance and frequency lines.
Same display
Backlit, with an analog bar graph, on both meters.
Same protection
CAT IV 600V and CAT III 1000V, UL, CSA and TÜV listed.
Same ownership
One 9 V cell, about 200 hours, and a lifetime warranty.
Verdict
Temperature work: 179. Purely electrical work: 177.
OhmSavvy hasn't hands-on tested either meter, so our editorial scores (4.7 for the 179, 4.6 for the 177) rest on Fluke's published specifications. Everything below comes from the two Fluke product pages, and the tolerances are our arithmetic on Fluke's numbers, not measurements.
The two accuracy tables, line by line
Fluke prints a short accuracy table on each 170-series page, giving the best accuracy, finest resolution and maximum for every function. Here are the 177 and 179 tables merged. The only row that differs is the last one.
| Function | Accuracy (both) | Best resolution | Maximum | 177 | 179 |
|---|---|---|---|---|---|
| DC volts | ±(0.09% + 2) | 0.1 mV | 1000 V | Yes | Yes |
| AC volts, True RMS | ±(1.0% + 3) | 0.1 mV | 1000 V | Yes | Yes |
| DC current | ±(1.0% + 3) | 0.01 mA | 10 A | Yes | Yes |
| AC current, True RMS | ±(1.5% + 3) | 0.01 mA | 10 A | Yes | Yes |
| Resistance | ±(0.9% + 1) | 0.1 Ω | 50 MΩ | Yes | Yes |
| Capacitance | ±(1.2% + 2) | 1 nF | 10,000 µF | Yes | Yes |
| Frequency | ±(0.1% + 1) | 0.01 Hz | 100 kHz | Yes | Yes |
| Temperature | ±(1.0% + 10) | 0.1 °C | −40 to 400 °C | No | Yes |
Fluke footnotes these as the best accuracy for each function, which usually means the most favorable range and frequency band. Individual ranges can be looser, so treat every worked figure below as the best case. The users manual holds the full per-range tables.
Worked tolerances for everyday readings
A spec of ±(a% + n) means a percent of the displayed value plus n steps of the last digit. On a 6,000-count meter the step depends on the range: 0.001 V on 6 V, 0.01 V on 60 V, 0.1 V on 600 V. The table below works that out for readings an electrician or technician takes every week. Because the tables match, each row holds for the 177 and the 179 alike.
| Reading | Spec | % part | Count part | Allowance |
|---|---|---|---|---|
| 5.000 V DC | 0.09% + 2 | 4.5 mV | 2 mV | ±6.5 mV |
| 24.00 V DC | 0.09% + 2 | 21.6 mV | 20 mV | ±41.6 mV |
| 120.0 V AC | 1.0% + 3 | 1.2 V | 0.3 V | ±1.5 V |
| 480.0 V AC | 1.0% + 3 | 4.8 V | 0.3 V | ±5.1 V |
| 20.00 mA DC loop | 1.0% + 3 | 0.20 mA | 0.03 mA | ±0.23 mA |
| 60.00 Hz | 0.1% + 1 | 0.06 Hz | 0.01 Hz | ±0.07 Hz |
Two things stand out. At low readings the count part can be as large as the percentage part: on 24 V, the two counts (20 mV) nearly equal 0.09% of the reading (21.6 mV). And AC volts are about ten times looser than DC on both meters, so a 120 V receptacle that reads anywhere from 118.5 V to 121.5 V on the display is inside the published tolerance if the true value is 120.0 V. For a 4–20 mA loop the ±0.23 mA allowance at full scale is just under 1.5% of the 16 mA span, which is fine for finding a dead transmitter but not for calibrating one. Our guide to measuring milliamps covers the series connection.
How the 179's temperature spec behaves
The temperature line, ±(1.0% + 10), has a large count term. At 0.1 °C resolution, ten counts are a fixed 1.0 °C, so near room temperature that fixed part is most of the allowance, and only at a few hundred degrees does the percentage part take over.
These are the meter's figures alone. Any thermocouple probe has its own tolerance, and the reading you see includes both, so the real uncertainty is wider than the bars. In practice that means the 179 is right for trend and difference work: supply versus return air, a bearing that runs warmer than its twin, a heat sink climbing under load. For a temperature you must certify to a fraction of a degree, use a dedicated thermometer. To check that a probe itself is sound, see how to check a thermocouple with a multimeter.
A difference reading cancels some of the error
Where Fluke's 179 page disagrees with itself
The comparison table further down Fluke's 179 page doesn't match its own spec table in three places. The spec table is more detailed and agrees with the 177 page, so we use it throughout.
| Item | 179 spec table | 179 comparison table |
|---|---|---|
| Capacitance maximum | 10,000 µF | 9,999 µF |
| Temperature top end | 400 °C | 400 °F (with −40 °C at the low end) |
| Small current | 0.01 mA best resolution | Lists µA as an extra function |
The current row matters most. If the 179's finest step is 0.01 mA (10 µA), it can't resolve a furnace flame-sensor signal of a few microamps, whatever the comparison table says. For that check, look at the Fluke 116 or 87V, discussed in our Fluke 87V vs 179 comparison.
Which features justify paying more for the 179?
Upgrades usually bundle several improvements, so buyers assume the dearer model is better across the board. This pair breaks that habit. Go through everything a buyer might hope the step up delivers, and only two lines change: the temperature function itself and the 80BK probe that ships with it. Every other item on the list is either identical on the 177 or missing from both meters.
| What you might expect from the step up | Fluke 179 vs 177 | Does it justify the extra? |
|---|---|---|
| Temperature, −40 °C to 400 °C | 179 only, ±(1.0% + 10) | Yes, if you take temperatures |
| A thermocouple probe in the box | 179 only, 80BK | Yes; it saves buying one |
| Tighter DC or AC volts accuracy | Same ±(0.09% + 2) and ±(1.0% + 3) | No |
| Finer current for flame sensors | Same 0.01 mA best resolution | No; neither resolves a few µA |
| Higher safety rating | Same CAT IV 600V / CAT III 1000V | No |
| Better display or backlight | Same 6,000 counts, bar graph, backlight | No |
| Longer battery life or warranty | Same 200 h on a 9 V; lifetime on both | No |
| LoZ or non-contact voltage | Missing on both | No |
That makes the value test unusually simple. Count how often a temperature reading would save you fetching another instrument. If it is part of most service calls, as on HVAC, motor and bearing checks, the 179 pays for itself in time and in the probe you don't have to buy separately. If you can't name a job in the last month where you needed one, the extra spend buys a dial position that will sit unused, and the 177 is the better value. Nothing about the 179 makes your volts or amps readings more trustworthy.
Which jobs need the 179's extra function
Because the electrical specs are identical, the right question is how often you'd reach for a temperature reading. If the answer is weekly, carrying the 179 saves a second instrument. If it is never, the 177 is the same meter for less.
| You are… | Better fit | Why |
|---|---|---|
| HVAC technician | Fluke 179 | Temperature splits and motor checks in the same meter as volts and capacitance |
| Industrial maintenance technician | Fluke 179 | Bearing, motor-frame and enclosure temperatures alongside electrical readings |
| Residential electrician | Fluke 177 | Temperature rarely needed; identical volts, amps and CAT IV rating |
| Electronics repair at a bench | Fluke 179, or 177 | The 179 spots hot components; the 177 if you own a thermal tool already |
| Solar or battery installer | Fluke 177 | 1000 V DC and 0.09% DC on both; temperature is a nice-to-have |
| Owner of a clamp meter with temperature | Fluke 177 | No reason to pay twice for the same function |
If neither temperature nor the tight DC spec appeals, step down: the Fluke 175 vs 177 comparison covers the cheapest 170-series meter. If your work is mostly building wiring, the Fluke 177 vs 117 page weighs LoZ and VoltAlert against the 170 series. HVAC buyers should also read our best multimeter for HVAC technicians guide.
Safety: keep the thermocouple off live parts
Choose the Fluke 179 if…
- ✓You check supply and return air, motor housings, bearings or heat sinks and want temperature in the meter you already carry
- ✓You want the 80BK temperature probe included rather than bought separately
- ✓You would rather own the top of the 170 series once than add a thermometer later
Choose the Fluke 177 if…
- ✓Your work is volts, amps, ohms, capacitance and frequency, and you never need a temperature reading
- ✓You already carry a separate thermometer or a clamp meter with temperature
- ✓You want every electrical spec of the 179 for less
The bottom line
The Fluke 177 and 179 share every published electrical accuracy figure, the same backlit display, CAT IV 600V / CAT III 1000V rating, 420 g body, battery life and lifetime warranty. The 179 adds one thing: temperature from −40 °C to 400 °C at ±(1.0% + 10), with an 80BK probe in the box. Buy the 179 for HVAC, motor and maintenance work where temperature comes up; buy the 177 if your readings are purely electrical.
Frequently asked questions
What is the difference between the Fluke 177 and 179?+
Temperature. The 179 reads −40 °C to 400 °C at ±(1.0% + 10) with 0.1 °C resolution and comes with an 80BK probe. Otherwise Fluke publishes the same accuracy for both meters on DC and AC volts, current, resistance, capacitance and frequency, and they share the same backlit display, CAT IV 600V rating and lifetime warranty.
Is the Fluke 179 more accurate than the 177?+
No. Fluke's product pages list identical best accuracies: ±(0.09% + 2) on DC volts, ±(1.0% + 3) on AC volts, ±(1.0% + 3) and ±(1.5% + 3) on DC and AC current, ±(0.9% + 1) on resistance, ±(1.2% + 2) on capacitance and ±(0.1% + 1) on frequency. The 179 only adds a temperature line.
Does the Fluke 179 come with a temperature probe?+
Yes. Fluke lists the 80BK temperature probe in the 179's box, along with test leads, an installed 9 V battery and a quick reference guide. Fluke's 179 page doesn't state the probe's own rating, so check its markings before measuring hot surfaces; a probe can be rated lower than the meter's 400 °C ceiling.
Can the Fluke 179 measure microamps for a flame sensor?+
Not usefully, by the spec table. Fluke gives 0.01 mA (10 µA) as the best current resolution, while flame-sensor signals are typically a few microamps. The 179 page's comparison table does list µA, which conflicts, so we follow the more detailed spec table. For flame sensors, use a meter with a 0.1 µA range such as the Fluke 116 or 87V.
How accurate is the Fluke 179 at room temperature?+
About ±1.25 °C at 25 °C from the meter alone, worked from Fluke's ±(1.0% + 10) spec: 1% of 25 °C is 0.25 °C, and ten counts at 0.1 °C add 1.0 °C. The probe's own tolerance adds to that. At 200 °C the meter allowance grows to ±3.0 °C, and at 400 °C to ±5.0 °C.
Is the Fluke 179 worth paying more for than the 177?+
Only if you will use temperature. The extra money buys one function, −40 °C to 400 °C at ±(1.0% + 10), plus an 80BK probe in the box. It does not buy better electrical accuracy, a higher safety rating, a longer warranty or finer current resolution, because Fluke lists those identically for both meters. Without temperature work, the 177 is the better value.
Should an HVAC technician buy the 177 or the 179?+
The 179, in most cases. HVAC work involves temperature splits, motor checks and capacitor tests, and the 179 handles all three with the probe included, at the same electrical accuracy as the 177. Pick the 177 only if you already carry a clamp meter or thermometer that covers temperature and want to avoid paying for it twice.
Related